GlucoSmart ® Plus Blood Glucose Measurement System |
||
![]() |
||
|
Model |
AGM-2200 | |
| Method | Electrochemical with GlucoSmart ® Plus blood glucose test strips | |
| Test strips | GlucoSmart ® Plus blood glucose test strips (PZN 4172778) | |
| Measurement Range | 20 - 900 mg/dl, bzw. 1 - 50 mmol/l | |
|
Units (changeable) |
[mg/dl], [mmol/l] when changing the unit, also the results in the memory will be recalculated |
|
|
Measurement Time |
10 seconds | |
| Calibration | Code is changed by buttons | |
| Energy | lithium battery (CR2032), DC 3 V | |
| Battery life time |
~1000 cycles (1 year when testing 3 times a
day) CR 2032 is the most spread battery type in small electronic devices. It can be bought in pharmacies, photo shops worldwide. |
|
| Size of the meter | 62,5mm width; 100,0mm length; 23,0mm height | |
| Weight |
61 g (with battery) |
|
| Type of Display |
LCD / 41,0mm x 41,5mm |
|
| Size of Display |
250 results with date, time and
|
|
| Data Storage |
up to 4 daily signals free adjustable |
|
| Calibration Method |
whole blood (YSI
2300 Stat Plus) |
|
| PC Interface | RS-232C | |
| Energy safe function | 3 minutes after the last user interaction | |
The unit was developed to be used within the following environmental conditions |
||
| Operating Temperature Range | 10-40 °C | |
| Relative Humidity | < 85% | |
| Altitude | up to 2500 m above NN | |
| Allowed Hematocrit | 20-60% | |
| Conducted Noise | installation category I (overvoltage category I) IEC664 | |
| Pollution Category | 2 to IEC664 | |
|
|
||
GlucoSmart ® Plus blood glucose test strips |
||
|
|
To use with | Exclusively with GlucoSmart ® Plus Blood Glucose Meter model AGM2200 |
| Electrode material | Gold | |
| Blood insertion method | capillary insertion | |
| Method | electrochemical | |
|
Necessary blood probe |
2 µl (minimum 1,5) | |
| Size | 33 mm x 7,5 mm | |
| Shelf life time | Usable within 4 months after the first opening of the original vial. | |
| Storage | In the original vial between 1 to 30°C. | |
| Packaging Unit | 50 test strips in 2 vials with drying agent with each 25 strips. | |
|
|
||